7.8 Replacement procedures
Introduction
This section gives instructions for the replacement procedures to be performed by the user of ÄKTA
pure.
WARNING
Disconnect power. Always disconnect power from the instrument before replacing
any component on the instrument, unless stated otherwise in the user documenta-
tion.
In this section
This section contains the following subsections:
See pageSection
3177.8.1 Replace tubing and connectors
3197.8.2 Replace internal tubing in Fraction collector F9-C
3227.8.3 Replace the Mixer
3237.8.4 Replace the O-ring inside the Mixer
3257.8.5 Replace the UV monitor U9-M flow cell
3287.8.6 Replace the UV monitor U9-L flow cell
3307.8.7 Replace Flow restrictor
3317.8.8 Replace the inlet filters
3327.8.9 Replace the pump head check valves
3357.8.10 Replace pump piston seals of Pump P9 or P9H
3447.8.11 Replace pump piston seals of Pump P9-S
3517.8.12 Replace pump pistons
3527.8.13 Replace pump rinsing system tubing
316 ÄKTA pure User Manual 29119969 AB
7 Maintenance
7.8 Replacement procedures