7.3.2 Replace the inline filter
Maintenance interval
Replace the inline filter that is located in the top section of the Mixer every week, or when required, for
example when the filter becomes clogged.
Required material
The following materials are required:
Online filter kit
Forceps
Gloves
Instruction
Follow the instructions below to replace the inline filter that is located in the top of the Mixer.
Use forceps and gloves during the replacement procedure to avoid contaminating the Mixer
components.
Tip:
ActionStep
Unscrew the top section of the Mixer.1
Remove the old filter using forceps. Replace the support net if this is damaged. Fit the
new filter.
2
Check the O-ring of the Mixer. If the O-ring is damaged, replace it according to Section 7.8.4
Replace the O-ring inside the Mixer, on page 323.
3
While holding the Mixer upright, screw the top section back onto the Mixer.4
266 ÄKTA pure User Manual 29119969 AB
7 Maintenance
7.3 Weekly maintenance
7.3.2 Replace the inline filter