9.8.12 System settings - Data collection
The table below describes the Data collection related system settings available for ÄKTA pure.
DescriptionInstruction name
The Data collection settings determine the maximum number of
data points collected for a given curve. Data reduction occurs if
the maximum number of data points is exceeded. To avoid data
reduction, set the maximum number of data points to be collected
to 180000 or insert a New Chromatogram instruction in the
method.
Note:
The default setting is 54000 data points, which corresponds to 1.5
h for a signal of 10 Hz.
Data collection
ÄKTA pure User Manual 29119969 AB 465
9 Reference information
9.8 System settings
9.8.12 System settings - Data collection