5.8.2 Prepare Fraction collector F9-R
Introduction
This chapter describes how to prepare and assemble Fraction collector F9-R before a run.
Fraction collector F9-R is connected to ÄKTA pure and controlled by UNICORN. Control of the fraction
collector can be achieved automatically in a method run, or manually.
Prepare the fraction collector
Before starting to prepare the Fraction collector F9-R, check the fractionation settings in the method
to be run. Perform the steps described below according to the settings in the method.
•
Assemble the Tube rack
•
Insert collection tubes
•
Adjust the Delivery arm
•
Update System Settings in UNICORN
Assemble the Tube rack
Illustration of the Tube rack
FunctionPart
Single cutout1
L-shaped cutout2
Bowl3
Tube support4
Tube guide5
Tube holder6
Note that the tube guide has both single and L-shaped cutouts, while the tube holder only has
single cutouts.
Note: