9.4 Recommended tubing kits for prepacked columns
This section specifies recommended tubing kits for prepacked columns used with ÄKTA pure.
Where alternative tubing kits are listed for a given column, the Pressure control option may
be used instead of changing the tubing kit to adapt to different run conditions. Refer to Sec-
tion 9.15 Pressure control, on page 501 for details.
Note:
CommentsRecommended Tubing KitsColumn name
Advanced system
2
Standard sys-
tem
1
0.5 with Pressure
control or 0.75
0.5All HiTrap 1 ml
(including HisTrap™, GSTrap™,
StrepTrap™)
0.75 with Pressure control for
maximum flow rate combined
with high viscosity.
0.750.5 or 0.75All HiTrap 5 ml
(including HisTrap, GSTrap except
4B, StrepTrap)
0.5 with Pressure
controlor 0.75
0.5 or 0.75GSTrap 4B, 5 ml
0.5 with Pressure
control or 0.75
0.5 or 0.75All HiScreen
0.50.5RESOURCE™ Q, S, ETH, ISO, PHE and
RPC 1 ml
0.50.5RESOURCE RPC 3 ml
0.75 with Pressure control for
maximum flow rate combined
with high viscosity.
0.750.5 or 0.75RESOURCE Q, S 6 ml
Avoid combination of column
valve V9-C and pH position in
valve V9-pH.
0.50.5SOURCE™ 15RPC ST 4.6/100
Resolution may be improved with
0.25.
Avoid combination of column
valve V9-C and pH position in
valve V9-pH.
0.50.5SOURCE 15Q. S, PHE 4.6/100 PE
Resolution may be improved with
0.25.
Use 0.25 for maximum resolution.
Avoid column valve V9-C and pH
position in V9-pH.
0.25 or 0.50.25 or 0.5Mini Q™, Mini S™ 4.6/50 PE
428 ÄKTA pure User Manual 29119969 AB
9 Reference information
9.4 Recommended tubing kits for prepacked columns