There are five main types of modules (named components in UNICORN) to select from:
Valves and pumps
Monitors and sensors
Fraction collectors
Other (e.g., I/O-box)
Core components (always present)
Multiple choices are not shown unless a component has been selected. The second component is only
available when the first has been selected. The selection made is reflected in which instructions and
phase properties that are available.
Edit system settings
It may be necessary to edit the System Settings when the configuration of the system is changed. For
example, if the change in configuration affects the delay volume following the UV monitor (or other
monitor connected via the I/O-box) the appropriate system settings for Tubing and delay volumes
have to be updated. This is to ensure that the fractions marked in the chromatogram corresponds to
the actual collected fractions.
Other system settings might also need to be edited for some optional modules.
The Edit dialog in which to edit the system settings is shown below.
In this dialog it is possible to set the parameters for the available instructions.
All system settings available for ÄKTA pure are found in Section 9.8 System settings, on page 449.
130 ÄKTA pure User Manual 29119969 AB
4 System configuration
4.1 Configuration overview