DescriptionInstruction name
Fraction collector lamp: Lamps in the fraction collector chamber
on or off.
Fraction collector lamp
The Peak fractionation parameters set the detection parameters
for peak collection, that is they decide when a peak starts and
ends. This information is used by the instructions Peak fractiona-
tion and Peak frac in outlet valve in order to start/end the peak
collection.
Peak fractionation parameters
Fraction collector F9-R
DescriptionInstruction name
Drop sync synchronises tube change to drop release. The available
settings are on or off. It is recommended to use Drop sync for
flow rates below 2 ml/min. Higher flow rates can however be used,
depending on the properties (for example viscosity) of the liquid.
Fractionation settings
Determines whether the fraction number is reset at the end of a
method or not.
Note:
The default setting is Reset.
Fractionation numbering mode
Drop sync on or off. It is recommended to use this setting for flow
rates below 2 ml/min. Higher flow rates can be however be used,
depending on the properties (e.g. viscosity) of the liquid.
Fractionation settings frac 2
Determines whether fraction number for the second fraction col-
lector is reset at the end of a method or not.
Note:
The default setting is Reset.
Fractionation numbering mode
frac 2
Peak fractionation parameters sets the detection parameters
for peak collection, i.e. it determines when a peak starts and ends.
This information is used by the instructions Peak fractionation,
Peak fractionation frac 2 and Peak frac in outlet valve in order
to start/end the peak collection.
Peak fractionation parameters
144 ÄKTA pure User Manual 29119969 AB
4 System configuration
4.2 Configure modules
4.2.11 Configuration of fraction collectors