ActionStep
Lift up the top section of the Mixer and pull away the old O-ring inside.4
Wet the new O-ring with 20% ethanol and fit it in position. Make sure that the inline filter
is still in position.
5
Reassemble the Mixer components and, while holding the Mixer upright, screw the top
section back onto the Mixer.
6
324 ÄKTA pure User Manual 29119969 AB
7 Maintenance
7.8 Replacement procedures
7.8.4 Replace the O-ring inside the Mixer