Possible cause and actionProblem
Air trapped in the pump heads
Partially blocked solvent filters
Leaking connections
Piston seal leakage
Check valve malfunction
Piston damaged
See Examples of pump pressure curves, on page390 for examples
of pump pressure curves.
Erratic pump pressure
The cable between the system pump and the ICU is not con-
nected.
Contact Service.
The cable between the sample pump and the ÄKTA pure instru-
ment is not connected.
Connect the cable.
A fuse in the instrument ICU is broken.
Contact Service.
The pump is not found by
the instrument
The air intake on the rear or on the left side of the instrument
is covered
Make sure that none of the air intakes on the instrument are
covered.
Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
The internal temperature of
the pump is too high
The pressure has increased due to increased viscosity
The viscosity increases in cold room. Lower the flow when per-
forming runs in a cold room.
High pressure alarm
388 ÄKTA pure User Manual 29119969 AB
8 Troubleshooting
8.6 Troubleshooting: Pumps

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AKTApure and is the answer not in the manual?

GE AKTApure Specifications

General IconGeneral
BrandGE
ModelAKTApure
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals