DescriptionInstruction name
Delay volume: Monitor to frac is used to define the delay volume
between the monitor and the Fraction collector. The instruction
is used to make sure that the collected fractions correspond to
the fractions indicated in the chromatogram. The instruction is
available only when the Fraction collector is selected in the com-
ponent list.
Delay volume: Monitor to frac
Delay volume: Monitor to frac 2 is used to define the delay volume
between the monitor and the second Fraction collector. The in-
struction is used to make sure that the collected fractions corre-
spond to the fractions indicated in the chromatogram. The instruc-
tion is available only when the second Fraction collector is selected
in the component list.
Delay volume: Monitor to frac 2
ÄKTA pure User Manual 29119969 AB 461
9 Reference information
9.8 System settings
9.8.8 System settings - Tubing and Delay volumes