E-PSM(P)W Specifications
7-3
Document no. 2044677-001
2 E-PSM(P)W Specifications
2.1 General specifications
Module size 55 x 135 (175 w/ tab) x 145 mm
W x D x H 2.2 x 5.3 (6.9 w/ tab) x 5.7 in
Module weight 0.6 kg /1.4 lb.
Operation temperature 5 to 40°C / 41 to 104°F
2.2 Typical performance
2.2.1 NIBP
NOTE: Non-invasive blood pressure measurement is intended for patients weighing over 5 kg
(11 lb.)
Oscillometric measurement principle.
Measurement range adult 25 to 260 mmHg
child 25 to 190 mmHg
infant 15 to 140 mmHg
Pulse rate range accepted 30 to 250 bpm
Typical measuring time adult < 30 s
infant < 25 s
Inflation limit adult 280 ±10 mmHg
child 200 ±10 mmHg
infant 150 ±5 mmHg
Venous stasis adult 80 ±5 mmHg / 2 min.
child 60 ±5 mmHg / 2 min.
infant 40 ±5 mmHg / 1 min.
Venous stasis pressure may be lower than the values above if the patient has low blood
pressure. The venous stasis pressure adapts to the measured mean pressure being the same
as mean pressure but always at least the following:
Infant 20 ± 5 mmHg
Child 30 ± 5 mmHg
Adult 40 ± 5 mmHg
2.2.2 ECG
Lead type 3-Lead (I, II, III)
5-Lead (I, II, III, aVR, aVL, aVF, V)
Sweep speeds 12.5, 25, 50 mm/sec.
Defibrillation protection 5000 V, 360 J
Recovery time 5 s
Input impedance >2.5 M (10 Hz)
Allowable offset ±1VDC
Gain range 0.2 to 5.0 cm/mV
Display filter
Diagnostic 0.05 to 150 Hz
Monitoring 0.5 to 30 Hz (-3 dB, with 50 Hz reject filter)
0.5 to 40 Hz (-3 dB, with 60 Hz reject filter)