Electrical Safety Tests
3-13
Document no. 2044677-001
(6) Set the power switch on the device to OFF.
(7) Repeat the steps in this procedure using the appropriate SPO2 Test Body. Connect the
SPO2 Test Body to the blue SPO2 connector of the device under test.
2.10.1 Acceptance criteria
If measured reading is greater than the appropriate specification below, the device under test
fails. Contact GE Technical Support.
• 50 µA at 120-240 VAC using the ECG cable.