E-PSM(P)W module introduction
1-29
Document no. 2044677-001
4 E-PSM(P)W module introduction
The E-PSMW and E-PSMPW modules provide general hemodynamic parameters.
Figure 15 Patient Side Module, E-PSMPW
Table 1 Options of Patient Side Module
4.1 Intended purpose (Indications for use)
The Patient side module E-PSM(P)W is indicated for monitoring of hemodynamic parameters of
all hospital patients. The hemodynamic parameters of the module comprise ECG (including
ST-Segment and arrhythmia), impedance respiration,
oscillometric NIBP (sys/dia/mean),
temperature, SpO
2 (including monitoring during conditions of clinical patient motion), and invasive
blood pressure. Impedance respiration measurement is indicated for patients ages three years
and up. The NIBP measurement is indicated for patients who weight 5kg (11 lb) or up.The
E-PSM(P)W is intended for all hospital departments including intra-hospital
transport but
excluding harsh physical environment like MRI.
4.2 Monitor software compatibility
B30 Patient Monitor using software L-DICU08
Parameter E-PSMPW E-PSMW
Two invasive blood pressures x
Impedance respiration x x
ECG x x
Pulse oximetry x x
Two temperatures x x
NIBP x x