2Specifications
2.1 General specifications
N-FCREC
Module size, W × D × H90 × 169 × 191 mm, 3.54 × 6.65 × 7.51 in
Module weight 1.0 kg/2.2 lb.
Power consumption 2.7 W standby, 11.5 W printing
N-FREC
Module size, W × D × H90 × 169 × 191 mm, 3.54 × 6.65 × 7.51 in
Module weight 0.8 kg/1.7 lb.
Power consumption 1.2 W standby, 10 W printing
N-FC
Module size, W × D × H46 × 169 × 191 mm, 1.81 × 6.65 × 7.51 in
Module weight 0.5 kg/1.1 lb.
Power consumption 1.5 W typical
2.1.1 Environmental specifications
Operating temperature +10...+40 °C (+50 to 104 °F)
Storage temperature -20...+60 °C (-4 to +140 °F)
Atmospheric pressure 666...1060 hPa /
(67...106 kPa)
(500...800 mmHg)
(666...1060 mbar)
Humidity 10...95% non-condensing (in airway 0...100%, condensing)
Protection against electrical shock Type BF
2.1.2 Functional alarms
Funtional alarms for
− Blocked sample line
− D-Fend replacement
− D-Fend check
2.2 CO
2
measurement
2.2.1 Typical performance
EtCO
2
End-tidal CO
2
concentration
FiCO
2
Inspired CO
2
concentration
Measurement range 0 to 20 vol% (0 to 20 kPa, 0 to 150 mmHg)
Accuracy
CO
2
concentration 0 to 15 vol% ±(0.2 vol% +2% of reading)
CO
2
concentration 15 to 20 vol% ±(0.7 vol% +2% of reading)
Measurement rise time < 300 ms with nominal flow
Adjustable low and high limits for EtCO
2
and FiCO
2
.