B30 Patient Monitor
2-7
Document no. 2044677-001
WARNING When detaching modules, be careful not to drop them. Always support with
one hand while pulling out with the other.
NOTE: Only one E-PSM(P)W module and one N-F(C)(REC) module can be attached to B30.
NOTE: If you want to install both modules, you must install the Extension Module, (N-F(C)(REC)
first and attach the E-PSM(P)W module to it.
1.6 Functional check
Refer to the 3. Functional check in Chapter 3 for procedure.
1.7 Visual indicators
Function Specification Explanation
External power supply Green LED Indicates when monitor is powered from
mains or ext. DC
Battery operation Green LED Indicates when monitor is powered from
internal batteries
Battery condition Orange LED Indicates when monitor is charging
batteries (solid) or battery failure (flashing).
Alarm LEDs Red LED Indicates a life threatening situation
Yellow LED Indicates serious but not life threatening
problems
Alarm Light Highly visible Red/
Yellow light Ease alarm detection from distance.
Brightness of the light and enabling the
alarm light function are user configurable.