Functional check
3-15
Document no. 2044677-001
−
Waveform 2 - P1
−
Waveform 3 - P2
Invasive Pressures - P1 ‘ART’ Setup - Label - ART
- P2 ‘CVP’ Setup - Label - CVP
Others -SPO2 Setup - Pleth Scale - AUTO
3.2.2 Display
1. Check that the picture on the screen is displayed properly.
"
3.2.3 Keyboard(s)
Tests with the Command Board:
− Press the
Monitor Setup key. Turn the ComWheel in both directions and check that the
cursor in the menu moves correspondingly. Select Normal Screen and check that the
menu disappears from the screen.
Check the rest of the menu keys by pressing them one by one.
"
3.2.4 Frame unit
1. Check that the clock on the screen shows correct time. Readjust the time and date, if
necessary.
"
3.2.5 Extension Module with CO
2
measurement
Wait until the message ‘Calibrating gas sensor’ disappears from the screen.
1. Block the tip of the sampling line with your finger and check that the message ‘Sample
line blocked’ appears on the monitor screen within 30 seconds.
2. Detach the Mini D-fend and check that the message ‘Check D-fend’ appears on the
monitor screen within 30 seconds.
Breathe to the sampling line briefly. Check that the CO2 information is updated on the
screen.
"
3.2.6 Multiparameter Hemodynamic Modules
ECG and RESP measurements
1. Connect an ECG cable to the module. Connect the cable leads to a patient simulator.
Check that all ECG and impedance respiration information is shown on the monitor
screen as configured on the simulator.
Turn the simulator off. Check that the ëAsystoleí and ëApneaí messages are displayed.