Electrical Safety Tests
3-11
Document no. 2044677-001
• GND (Earth) - CLOSED
(10) Read and record the current leakage indicated on the tester.
(11) Change leakage tester switches to:
• Polarity - NORMAL
• Neutral - OPEN
• GND (Earth) - CLOSED
(12) Read and record the current leakage indicated on the tester.
(13) Change leakage tester switches to:
• Polarity - NORMAL
• Neutral - OPEN
• GND (Earth) - CLOSED
(14) Read and record the current leakage indicated on the tester.
(15) Set the power switch of the device to OFF.
(16) Repeat the steps in this procedure using the appropriate SPO2 Test Body. Connect the
SPO2 Test Body to the blue SPO2 connector of the device under test.
2.9.1 Acceptance criteria NC
With Ground and Neutral CLOSED - If reading is greater than 10 µA, the device under test fails.
Contact GE Technical Support
2.9.2 Acceptance criteria SFC - ground (earth), line or neutral open
If any reading is greater than 50 µA, the device under test fails. Contact GE Technical Support.