GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
3-46 Section 3-8 - System Configuration
3-8-10 Technical Support Tab
The System Configuration Technical Support tab is used when new hardware has been installed, or
when it necessary to enable new functionality.
Monitoring -
should always be enabled. (this provides constant monitoring of critical
parameters to ensure safe operation of the system).
Watchdog - normally should not be enabled. When enabled, if there is a system lock-up
caused by the Back End, this will turn OFF power at the Front End.
MOD Fan - should be enabled only on BEPs based on Pentium IV - this provides faster
operation of the MOD. Enabling this option in the wrong type of BEP may
cause damage to the MOD or MO media.
Use Front End - normally should be enabled.
iLinq - should be enabled only on systems that use the iLinq option.
Figure 3-28 Technical Support Tab
RFI-configured systems
RFT-configured systems