GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
8-22 Section 8-2 - Cover Replacement Procedures
8-2-11-4 Control Console Top Cover Installation Procedure (Upper Section)
1) Raise the control console to its maximum height.
2) Return the top cover to its original position and fasten it with six of its eight screws using a short
screw driver. The remaining two screws are located next to the location of the AC/VGA connectors
box, and can be accessed from the opened space at the rear of the gas spring using a long screw
driver.
3) Mount the AC/VGA connectors box into its original position and fasten it with its two screws. These
screws can be accessed from the opened space at the rear of the gas spring using a long screw
driver, as shown in Figure 8-19 "Removing Screws from the AC/VGA Connectors Box" on page 8-
18.
4) Lift the front section of the gas spring cover from the top cover collar and fasten it with its three
screws, as shown in Figure 8-14 "Gas Spring Cover Screws" on page 8-12.
5) Return the rear section of the gas spring cover into its original position and fasten it with its two
screws.
6) Install the monitor, as described in 8-3-1 "Vivid 3N Monitor 15" Replacement - Procedure" on page
8-28.
7) Plug the AC cable and VGA cable to the rear of the control console, as shown in Figure 8-18 "AC
and VGA Cables at Rear of Control Console" on page 8-17.
8) Return all peripherals to their original locations.