GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 8 Replacement Procedures 8-9
8-2-6 Top Cover (Lower Section) Replacement Procedures
8-2-6-1 Tools
Use the appropriate Phillips-type screw drivers, and a wire cutter as indicated in the replacement
procedures for the lower section of the top cover.
8-2-6-2 Preparations
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in Chapter 3 - Installation.
8-2-6-3 Top Cover Lower Section Removal Procedure
1) Remove the side covers, as described in 8-2-2 "Side Covers Replacement Procedures" on page 8-
4.
2) Remove the rear cover, as described in 8-2-4 "Rear Cover Replacement Procedures" on page 8-7.
3) Raise the control console to its maximum height.
4) Remove the two screws located in the brackets underneath the lower section of the top cover, one
on either side of the unit, as shown below:
Figure 8-9 Top Cover (Lower Section) Screws
5) Remove the connectors panels cover, as described the 8-2-5 "Connector Panels Cover
Replacement Procedures" on page 8-8.
6) Unplug all the peripheral cables connected to the rear connector panels.
NOTE: Pay attention to the location of each cable, as each one must be reconnected to its
original location during the installation process.
Screws