GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
8-50 Section 8-4 - Front End Parts Replacement
8-4-4 TX Power Supply Replacement Procedure
8-4-4-1 Tools
Use the appropriate flat and Phillips-type screw drivers, a wire cutter, and the appropriate Hex wrenches
as indicated in the Tx power supply replacement procedure.
8-4-4-2 Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in Chapter 3 - Installation.
8-4-4-3 TX Power Supply Removal Procedure
1) Remove the right cover, as described in 8-2-2-3 "Side Covers Removal Procedure" on page 8-4.
2) Remove the DC power supply - refer to the 8-4-3-3 "DC Power Supply Removal Procedure" on
page 8-49.
3) Remove the two screws located at the top of the Tx power supply unit.
4) Remove the Tx power supply unit by pulling it out by the handle on the front panel.
8-4-4-4 TX Power Supply Installation Procedure
1) Return the Tx power supply unit to its original position.
2) Fasten the Tx power supply unit with the two screws at its top.
3) Install the DC power supply, as described in 8-4-3-4 "DC Power Supply Installation Procedure" on
page 8-49.
4) Reconnect the Tx power supply input cable.
5) Return the right side cover to its original position, as described in 8-2-2-4 "Side Covers Installation
Procedure" on page 8-4.
NOTE: When you install the Tx power supply unit, ensure that it slides into its guiding tracks.