IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
8-146 Section 8-8 - Peripherals
8-8-5-3 Preparations
1) Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panels Cover as described in 8-2-5-3 "Connector Panels Cover Removal
Procedure" on page 8-8.
3.) Unpack the Color Video Printer.
4.) Set the NTSC/PAL selector switch to PAL or NTSC as shown in Figure 8-130 "Color Video Printer
Window Display (INPUT SETUP top menu displayed)" on page 8-143.
5.) Install the Color Video Printer accessories, as described in the Operating Instructions supplied with
the printer.
8-8-5-4 Sony UP-21MD Color Video Printer Installation - Procedure1
NOTE: The Installation Procedure is for units that are equipped with a previously-installed B/W printer and/or
VCR. The color printer option includes a shelf with fastening belt, on which the printer is placed.
NOTE: If a B/W Printer and VCR are not installed, then the Color Video Printer can be placed on the top surface
of the Vivid 3N ultrasound unit (below the keyboard) and secured with the belt.
1.) Raise the control console to its maximum height.
2.) If a B/W Printer and/or a VCR are installed, disconnect all the cables and remove the units.
3.) Remove the Metal Bottom Cover, beneath the front part of the control console, as described in 8-
2-10-3 "Control Console Bottom Cover (Upper Section) Removal Procedure" on page 8-17.
4.) Remove the Compartment Tray as described in 8-6-7-3 "Gas Spring Removal Procedure" on page
8-105.
5.) Remove the Compartment Base as described in 8-2-10-3 "Control Console Bottom Cover (Upper
Section) Removal Procedure" on page 8-17.
6.) Route the Color Printer cables, found in the compartment, and let them drop down through the
opening left after the Metal Bottom Cover was removed (see step 3, above).
7.) Re-install the Metal Bottom Cover, as described in 8-2-10-4 "Control Console Bottom Cover (Upper
Section) Installation Procedure" on page 8-19.