3-42
Performance Tests
PDH Frame Analysis (Options UKK, [USB])
Error Detection
5. Use the Frame Generator 139264 Kbit/s and keys to select
and enable BIT ERROR.
6. Check that Errors and Remote alarm Led’s are ON.
7. Check that the HP 37717C counts Frame (FAS) Errors. The Code Error count
should be 0.
8. Press the Frame Generator key and check that the FRAME (FAS) and
CODE error count is 0 and the Alarm Led’s are OFF.
Alarm Detection
9. Set the Frame Generator SB parameter to [1000].
10. Check that the HP 37717C Remote Alarm Led is ON. Check that the Error Count
and Error Ratio is 0.
11. Reset the Frame Generator SB parameter to [0000].
FAS Error and AIS
12. Use the Frame Generator 139264 Kbit/s and keys to select
and enable FAS ERROR.
13. Check that the HP 37717C Frame Loss alarm Led is ON.
14. Press the Frame Generator 139264 Kbit/s key and check that the
FRAME (FAS) Error Count is 0 and all Alarm Led’s are now OFF.
15. Use the Frame Generator 139264 Kbit/s and keys to select
and enable AIS OUT.
16. Check that the HP 37717C Frame Loss and AIS alarm Led’s are ON.
17. Press the Frame Generator 139264 Kbit/s key and check that the
FRAME (FAS) Error Count is 0 and all Alarm Led’s are now OFF.
18. Press on the HP 37717C once to stop the measurement.
19. Disconnect the test equipment.
SELECT
ON/OFF
ON/OFF
SELECT
ON/OFF
ON/OFF
SELECT
ON/OFF
ON/OFF
RUN/STOP

Table of Contents

Other manuals for HP 37717C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 37717C and is the answer not in the manual?

HP 37717C Specifications

General IconGeneral
BrandHP
Model37717C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals