3-200
Performance Tests
ETSI/ANSI Receiver Monitor Levels (Option UKZ)
DS1 26dB Monitor
8. Connect a fixed 6dB Attenuator in the Additional Attenuation position of the
circuit of Figure 3-57.
9. Set the HP 37717C MONITOR GAIN to 26dB mode.
10. Press RESULTS [TROUBLE SCAN] , then RUN/STOP key to start the
measurement.
11. Ensure that NO TROUBLE is displayed on the RESULTS page after 30 seconds.
12. Press RUN/STOP key to stop the measurement.
DS1 30dB Monitor
13. Connect a fixed 6dB Attenuator and a fixed 3dB Attenuator in the Additional
Attenuation position of the circuit of Figure 3-57.
14. Set the HP 37717C MONITOR GAIN to 30dB mode.
15. Press RESULTS [TROUBLE SCAN] , then RUN/STOP key to start the
measurement.
16. Ensure that NO TROUBLE is displayed on the RESULTS page after 30 seconds.
17. Press RUN/STOP key to stop the measurement.