3-104
Performance Tests
STM-1/STM-4, 1550nm Optical Interface (Option URU)
STM-1/STM-4, 1550nm Optical Interface (Option URU)
NOTE An SDH Option must be fitted for this option to operate - see Appendix B.
Specification (Transmitter)
Specification (Receiver)
Description
Minimum and maximum Optical Rx Input levels are verified by looping the Optical
Tx and Rx ports through the Optical Attenuator and checking for error-free
operation with identical Tx and Rx patterns selected. The signal level to the Optical
Rx input is measured using the Power Meter and adjusted using the Optical
Attenuator for the required minimum and maximum levels.
Wavelength 1520 nm to 1565 nm
Power Output -1 dBm (Nominal)
Line Coding NRZ
Safety Class Class 1, FDA 21 CFR Ch.1 1040.10 and EN 60825
Data Rate 155.52 Mb/s or 622.08 Mb/s (Nominal)
Wavelength 1200 nm to 1600 nm
Maximum Input Power -8 dBm (for BER of 1.0E
-10
)
Line Coding NRZ
Sensitivity -26 dBm Minimum (wavelength=1550nm,
Modulation=100%,Data=2
23
-1, BER=1.0E
-10
)
PMP Electrical Input 150mV (Nominal)
PMP Impedance Nominal 50

Table of Contents

Other manuals for HP 37717C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 37717C and is the answer not in the manual?

HP 37717C Specifications

General IconGeneral
BrandHP
Model37717C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals