3-98
Performance Tests
STM-1 Optical Interface (Option UH1)
STM-1 Optical Interface (Option UH1)
NOTE An SDH Option must be fitted for this option to operate - see Appendix B.
Specification (Transmitter)
Specification (Receiver)
Description
The 155 Mb/s output of the optical transmitter is checked for power level.
The receiver sensitivity is verified by attenuating the transmitter output and
checking for no errors in a back to back mode.
Wavelength 1280 nm to 1330 nm
Fiber Power Output -9 dBm nominal
Line Coding NRZ
Safety Class Class 1, FDA 21 CFR Ch.1 1040.10 and EN 60825
Data Rate 155.52 Mb/s (Nominal)
Wavelength 1200 nm to 1600 nm
Maximum Input Power -8 dBm (for BER of 1.0E
-10
)
Line Coding NRZ
Dynamic Range 20 dB minimum
Sensitivity -28 dBm Minimum (wavelength=1300nm,
Modulation=100%,Data=2
23
-1, BER=1.0E
-10
)

Table of Contents