3-146
Performance Tests
Transmitted Jitter/Wander Amplitude Accuracy (Option A3K [A3Q])
5. Connect up the equipment as shown in Figure 3-51.
Figure 3-51 External Jitter Generation
6. Increase the Synthesizer output level until the first Bessel Null is observed on the
Spectrum Analyzer.
7. Check that the amplitude of the Oscilloscope waveform is between 1.76V pk_pk
and 2.84V pk_pk.
Wander Generation
1. Select and setup the Test
HP 37717C display as shown opposite.
TRANSMIT