MTS 793 Controller Hardware FlexTest GT Controller Connections
105
D/A Connections Digital-to-analog mezzanine cards support
up to six program or readout signals to
external devices. Each D/A signal is an
analog output within ±10 volts.
• Each readout signal is from a 16 bit
digital to analog converter.
• Each output is a ±10 V analog output.
• The Model 493.46 D/A mezzanine card
provides analog signals to the Model
493.76 Analog Out transition module.
Analog outputs are available at the
BNC connectors of a Model 493.76
Analog Out transition module for
external devices (DVM, oscilloscope).
+
-
+
-
+
-
Output 1
Output 2
Output 3
Output 4
Output 5
Output 6
+
-
+
-
+
-
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
J4 - J7
From
D/A
To
External
Device