MTS 793 Controller Hardware
FlexTest GT Controller Connections
78
• Valve connections require a valve driver mezzanine card be
installed in the I/O Carrier module. The following valve drivers
can be installed:
– Model 493.14 Valve Driver
– Model 493.15 3-Stage Valve Driver
• Analog I/O connections require an A/D or D/A mezzanine card be
installed in the I/O Carrier module. The following analog
mezzanine cards can be installed:
– Model 493.45 A/D Converter
– Model 493.46 D/A Converter
Sensor Connections
Sensor connections require a conditioner mezzanine card be installed
in the I/O Carrier module. The following conditioners can be installed:
Note Model 493.21B Digital Universal Conditioners are not compatible
with standalone FlexTest SE Controllers.
• Model 493.21B Digital Universal Conditioner
• Model 493.21BC Digital Universal Conditioner
• Model 493.25 Digital Universal Conditioner
• Model 493.47 Encoder
• Model 493.48 Acceleration Conditioner
Each Carrier module can include up to four mezzanine cards. Each
installed mezzanine card is assigned a specific I/O Carrier module front
panel connector (J4–J7 I/O). These connections can be used for any
type of sensor (provided the appropriate mezzanine card is installed).
A hardware interface file (.hwi) defines each type of module (and their
associated mezzanine cards) and maps each module location for the
system software. The hwi file and the physical locations for each type
of module and associated mezzanine cards must match.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com