EasyManuals Logo
Home>MTS Systems>Controller>FlexTest GT

MTS Systems FlexTest GT User Manual

Default Icon
240 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
MTS 793 Controller Hardware FlexTest GT Controller Connections
95
Transducer Identification Modules
Transducer ID modules (located at the end of the sensor cable) are
optional components that can store sensor calibration information just
like a sensor calibration file. Since the calibration information stays
with the sensors, transducer ID modules make it easier to change
sensors.
Calibration information can be stored in the transducer ID EEPROM or
in a sensor calibration file. When you load a station that does not have
a sensor file assigned to one or more of its signals, by default, the
system software looks for calibration information in the transducer ID.
The molded ID module has a removable cover and includes provisions
for:
A shunt calibration resistor
Up to three bridge completion resistors
A transducer ID circuit card
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest GT and is the answer not in the manual?

MTS Systems FlexTest GT Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest GT
CategoryController
LanguageEnglish

Related product manuals