MTS 793 Controller Hardware
FlexTest GT Controller Connections
90
Sensor Cables
A typical system is provided with a standard set of sensor cables as
specified on the System Cable/Jumper Plug 493 Package Selection
drawing (PN 700-000-656). Refer to this drawing for the most current
part numbers. Standard sensor cables do not have ID modules.
Sensor cables with an ID module built into them are available as an
option. Sensor cables with ID modules use a system cable to bridge
the sensor cable to the appropriate connector on the rear panel of the
chassis.
Sensor cable specifications
15 contact type D male EMI connector
Backshell–EMI metallized plastic
Cable type–For all applications including CE EMC Compliance,
use MTS cable material PN 050-530-101 (two 22 AWG twisted pairs
with inner and outer braided shield for the excitation and
feedback signals, plus a twisted quad for the ID signals, and a
single wire for the shunt calibration signal).
If purchased as an option, each sensor includes one cable with a
transducer ID module and a system extension cable. A system
cable is required between the transducer ID cable and the
controller’s chassis.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest GT and is the answer not in the manual?

MTS Systems FlexTest GT Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest GT
CategoryController
LanguageEnglish

Related product manuals