TDS 200-Series Digital Oscilloscope User Manual
91
Appendix A: Specifications
All specifications apply to the TDS 200-Series Digital Oscilloscopes
and a P2100 probe with the Attenuation switch set to 10X unless
noted otherwise. To meet specifications, two conditions must first be
met:
H The instrument must have been operating continuously for twenty
minutes within the specified operating temperature.
H You must perform the Self Cal operation, accessible through the
Utility menu, if the operating temperature changes by more than
5° C.
All specifications are guaranteed unless noted “typical.”
Specifications
Acquisition
Acquisition Modes Sample, Peak Detect, and Average
Acquisition Rate,
typical
Up to 180 waveforms per second, per channel (Sample
acquisition mode, no measurements)
Single Sequence Acquisition Mode Acquisition Stops After
Sample, Peak Detect Single acquisition, all channels
simultaneously
Average N acquisitions, all channels
simultaneously, N is selectable
from 4, 16, 64, and 128
Inputs
Input Coupling DC, AC, or GND
Input Impedance, DC
Coupled, all channels
1 MW ±2% in parallel with 20 pF ±3 pF