4
REMOTE INTERFACES
4-4 TP04300 Series Interface & Applications Manual
Section B:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Enable/Disable Version 1 Software
Refer to Chapter 3 to determine current software version.
Enable Version 1
Software
To enable prior Software Version 1 set "Compatible4000=FALSE" flag, as follows:
CAUTION
When editing the "Compatible4000" setting in \X-Stream.ini, do not change any other settings:
doing so can cause the System to not function, or to not function properly.
Step Action
1 If System is running, then press "Off" on Statusbar Display to exit
ThermoStream Control Software (TSCS). If needed, see Chapter 3 for
shutdown information.
2 Attach a keyboard, and a mouse, (and, preferably a monitor) to OCM front
panel ports. If needed, see Chapter 3 for keyboard, mouse, and monitor
connections.
3 Press front panel On switch to energize system.
4 Microsoft Windows NT operating system begins to boot
5 When Windows "Desktop" appears, then press "Start" key (or left-click "Start"
button in lower left, in Task Bar displayed at screen bottom)
6If TP04300A boots completely, then press "Start" key on keyboard, and
continue.
7 In START menu, navigate to PROGRAM, then ACCESSORIES menus, and
select (launch) either NOTEPAD or WORDPAD program.
8 In Notepad (or Wordpad) go to FILE, then OPEN, and set drop down menu,
"Files of Type," to "All Files"
9 Navigate to, and open the file: "D:\X-Stream\X-Stream.ini
10 In \X-Stream.ini file, edit "Compatible4000" flag to read
Compatible4000=FALSE (use uppercase letters in FALSE flag)
11 Press FILE, then SAVE, to save \X-Stream.ini.
12 Press FILE, then OPEN, then select and double click \X-Stream.exe to boot the
ThermoStream Control Software (TSCS) operating system.
13 Close NotePad (or WordPad): press FILE, then EXIT.
14 If the TSCS is already running, then press "Off" on Statusbar Display to exit the
TSCS, and do a normal system startup to reboot and install the new values.
15 The System is now running Version 1 commands.