ROUTINE MAINTENANCE
Sensor Calibration: Type T, Type K Thermocouples (DUT Mode)
TP04300 Series Interface & Applications Manual 5-25
Sensor Calibration: Type T, Type K Thermocouples (DUT Mode)
Introduction to Type
T and Type K
Thermocouple
Calibration
The sensor calibration procedures for Type T and Type K Thermocouples are identical.
There are 2 methods for calibrating the sensors:
Calibrator
External Sensor
Calibrator is the recommended method. This method uses an external voltage source (or
Calibrator) as the primary reference. The Calibrator inputs millivolt values for high and low set
points.
External Sensor is used for unique applications. It is recommended that you contact a
Temptronic representative before calibrating with this method.
This method uses an external temperature sensor as a primary reference.
Once the Sensor has been chosen from the Calibration Select Sensor Screen, the following
dialog box will appear:
1980_503.jpg
Calibrator Method
(Type T, Type K)
1980_504.jpg

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Temptronic ThermoStream TP04300 Series and is the answer not in the manual?

Temptronic ThermoStream TP04300 Series Specifications

General IconGeneral
BrandTemptronic
ModelThermoStream TP04300 Series
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals