ROUTINE MAINTENANCE
Drain Moisture from Pneumatics Filter Elements
TP04300 Series Interface & Applications Manual 5-7
4 Optional: spigot has barb fitting to attach rubber hose; put free end of hose in
drainage container
5 Grasp spigot knob, turn clockwise 1/4 turn to open spigot and drain condensate
6 Drain until no more fluid comes out
7 Close spigot: turn counter-clockwise 1/4 turn
8 If drained condensate is viscous (thick) or highly contaminated with particles,
then remove bowls and wipe them clean.