EasyManuals Logo
Home>Temptronic>Industrial Equipment>ThermoStream TP04300 Series

Temptronic ThermoStream TP04300 Series User Manual

Temptronic ThermoStream TP04300 Series
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #207 background imageLoading...
Page #207 background image
. . . . .
ROUTINE MAINTENANCE
Sensor Calibration: Type T, Type K Thermocouples (DUT Mode)
TP04300 Series Interface & Applications Manual 5-27
ATTENTION
The System must initially be calibrated with the Calibrator method before the External
Sensor method can be used.
Step Action
1 Run the system for half an hour at Ambient to stabilize components at normal
operating temperatures.
2 Toggle Head into Up position so System is not supplying main air flow
controlled to a setpoint (other than "Trickle" air flow).
NOTE: 4300B Systems should have the “Head: Lock” on. The “Head: Lock”
can be toggled on/off in the Utilities Screen.
3 Using a “Y” cable, connect a sensor from the System’s rear I/O port and from
an External Calibrator to the Main Air Thermal Cap:
1980_517.jpg
4 Once the sensor has been properly connected, press “NEXT >”.
5 Enter the desired “Low Calibration” setpoint (between -60.0 and 0.0) and press
“OK”.
Once “OK” is pressed, the System will ramp to the “Low Calibration” setpoint.
1980_510.jpg
Diagram:
Y Cable Setup

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Temptronic ThermoStream TP04300 Series and is the answer not in the manual?

Temptronic ThermoStream TP04300 Series Specifications

General IconGeneral
BrandTemptronic
ModelThermoStream TP04300 Series
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals