EasyManua.ls Logo

Feasa 10-F User Manual

Default Icon
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
FEASA LED ANALYSER
FUNCTIONAL VERSION
Get Data Mode
getxyi## - Return the xy Chromaticity & Relative Intensity values
Transmit Receive
getxyi##
getxyiall
0.xxxx 0.yyyy IIIII
Where:
## represents the Fiber Number and is a number in the range 01 – 20.
all represents the 20 Fibers in a Feasa 20F, or 3 Fibers in a 3F.
0.xxxx represents the x Chromaticity value
0.yyyy represents the y Chromaticity value
IIIII represents the Relative Intensity value
Description
This command is used to return the xy Chromaticity and Relative Intensity values for the LED
under the Fiber number ##. This command is used for testing White LED's.
This command should be preceded by a capture command to ensure valid LED data is stored
in the memory of the LED Analyser. The data for each LED can be read out one-at-the-time,
in any order. Note: it is important for the PC to use 01, 02, 03, etc to indicate Fibers 1, 2
and 3, etc.
Alternatively use the getxyiall command to read back the LED data for all 20 leds in
one command.
Example:
The PC transmits getxyi01 to the LED Analyser to instruct it to send the stored xy
Chromaticity and Relative Intensity data for the LED positioned under Fiber No 1. The LED
Analyser will return a string 0.xxxx 0.yyyy 12345 to the PC.
getxyi01
0.6461 0.3436 03456
Back to Inde x
28

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Feasa 10-F and is the answer not in the manual?

Feasa 10-F Specifications

General IconGeneral
BrandFeasa
Model10-F
CategoryMeasuring Instruments
LanguageEnglish

Summary

Important Information for Programmers

Colour and Saturation

Intensity

Setting Tolerance Limits

Capture Commands

AUTO CAPTURE - Store LED Data

Captures LED data automatically based on intensity.

MANUAL CAPTURE - Store LED Data

Captures LED data using a specific range.

AUTO CAPTUREPWM - Store PWM LED Data

Captures data from PWM LEDs automatically.

MANUAL CAPTUREPWM - Store PWM LED Data

Captures PWM LED data using a specific range.

Get Data Commands

getRGBI## - Get RGB and Intensity

Retrieves RGB colour components and intensity.

getHSI## - Get Hue, Saturation and Intensity

Retrieves Hue, Saturation, and Intensity values.

getINTENSITY## - Get Relative Intensity

Retrieves the relative intensity of the LED.

General Set/Put Commands

User Calibration Mode

UserCal Manual Procedure

Step-by-step guide for manual user calibration.

External Trigger Mode

Sequence Capture Mode

Capture Sequence Procedure

Outlines the steps to perform a sequence capture.

Step-by-Step Testing Guides

Step-by-Step LED Testing

Standard procedure for testing LED colour and intensity.

Specifications

Related product manuals