FEASA LED ANALYSER
FUNCTIONAL VERSION
User Calibration Mode
ResetUserCal##- Reset the standard usercal setting to factory default
Transmit Receive
resetusercal## OK
Where: ## is the calibration set number
Description
This command will reset the calibrationset back to factory default values. The Led analyser is
capable of storing fifty one (00-50) separate user Calibration sets on the board. Each one must
be reset separately.
The command resetusercal## resets the selected Calibration set to factory default.
The command resets:
- Absolute intensity factors to 0
- Relative intensity gains to 100
- xy and dominant wavelength offsets to 0
- User calibration date to 00/00/0000
- Reset RGB mode if selected and come back to standard mode
- Exposure factor to 1
- Log mode
of all the fibers in the user calibration set selected.
Example:
If you want to reset calibration set No.25
setcalibration25
OK
resetusercal
OK
The command GETCALIBRATIONSET will return the number of which stored Calibration set of
values are currently selected.
Example:
calibrationset03
Back to Index