EasyManuals Logo
Home>Feasa>Measuring Instruments>10-F

Feasa 10-F User Manual

Default Icon
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
FEASA LED ANALYSER
FUNCTIONAL VERSION
User Calibration Mode
getCALIBRATIONset - Select the Calibration Set to use
Transmit Receive
getcalibrationset ##
Where:
## represents the set value (## can be 00 to 50).
Description
This command will retrieve which calibration set you are accessing. The Led analyser is
capable of storing fifty one (51) separate user Calibration sets on the board.
The command SETCALIBRATION## stores the Calibration sets both on RAM and on EEPROM
(stays after power cycle but limits the number of set command operations to ~ 100K).
The command PUTCALIBRATION## only stores them on RAM and the analyser will default
back to what values were stored in the EEPROM after a power cycle.
The default calibration set is 00
Example:
If you want to know what set is active
getcalibrationset
03
Back to Index
69

Table of Contents

Questions and Answers:

Feasa 10-F Specifications

General IconGeneral
BrandFeasa
Model10-F
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals