FEASA LED ANALYSER
FUNCTIONAL VERSION
Get Data Mode
setPHOTOPIC - Set which Photopic Mode you require
Transmit Receive
setphotopic# OK
Where #: 0 is Photopic Response OFF
1 is Response Photometric (Default from the Factory)
2 is Response Radiometric
Description
The Led Analyser is shipped from the factory in photopic mode 1. This corrects the Absolute
Intensity to reproduce the photometric response (for lm, mcd). This can be verified by typing
the command getphotopic with the answer 1.
The photopic mode can be set to a radiometric response by sending the command
setphotopic2. This corrects the Absolute Intensity to reproduce the radiometric response (for
mW)
The photopic mode can be turned OFF by sending the command setphotopic0. With the
Analyser in this mode there is no effect on Absolute Intensity.
This command is only available on Software Rev F210 upwards