FEASA LED ANALYSER
FUNCTIONAL VERSION
Get Data Mode
getABSINT@@ Get the Absolute Intensity Value of the Led under test.
Transmit Receive
getabsint@@
#.####e±##
Where:
@@ represents the Fiber Number and is a number in the range 01 – 20.
all represents the 20 Fibers in a Feasa 20F, or 3 Fibers in a 3F.
#.####e±## represents the Intensity reading
Description:
This command will allow the user to interrogate the Led Analyser to return the absolute
Intensity Value for the Led under test at Fiber@@.
This command should only be used in conjunction with the Feasa LED Spectrometer.
Absolute Intensity values must first be measured with the Spectrometer. The LED Analyser
must then be calibrated with these values using the usercal program supplied with the
Analyser.
Alternatively use the getabsintall command to read back the LED data for all 20 leds
in one command.
Example:
Place fiber 01 over the Led to be tested. Using any of the Capture commands, capture the
data for that Led. Type the command getabsint01 and the analyser will respond with with
the intensity reading in the format #.####e±##
getabsint01
2.5000e-02
Back to Index
36