FEASA LED ANALYSER
FUNCTIONAL VERSION
Get Data Mode
getWSI## - Get the Dominant Wavelength, Saturation & Intensity
Transmit Receive
getwsi##
getwsiall
XXX SSS YYYYY
Where:
## represents the Fiber Number and is a number in the range 1 – 20.
all represents the 20 Fibers in a Feasa 20F, or 3 Fibers in a 3F.
XXX represents the dominant wavelength of the LED in nanometers.
SSS represents the Saturation (whiteness) and is a number in the range 0–100.
YYYYY represents the intensity of the LED.
Description
This command is used to get the value of the Dominant Wavelength for the LED under the
Fiber number ##. This command should be preceded by the capture C command to ensure
valid LED data is stored in the memory of the LED Analyser. The C1-C5 commands are
available but are not as accurate. The data for each LED can be read out one-at-the-time, in
any order.
Alternatively use the getwsiall command to read back the LED data for all 20 leds in
one command.
Example:
The PC transmits getwsi01 to the LED Analyser to instruct it to send the stored dominant
wavelength & Intensity for the LED positioned under Fiber No 1. The LED Analyser will return
a string xxx yyyyy to the PC.
getwsi01
513 100 12345
Back to Index
33

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Feasa 10-F and is the answer not in the manual?

Feasa 10-F Specifications

General IconGeneral
BrandFeasa
Model10-F
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals