FEASA LED ANALYSER
FUNCTIONAL VERSION
General set/put Commands
setBAUD - Change the baud rate
Transmit Receive
Setbaudxxxxx OK
Where: X = 9600, 19200, 38400, 57600, 115200 for Serial RS232 Port
Where: X = 9600, 19200, 38400, 57600, 115200, 230400, 460800, 921600 for USB Port
Description
This command will change the baud rate of the Serial and USB Port's in the Analyser.
The default Port settings of the Analyser are 57,600, 8 Data bits, 1 Stop bit and No Parity.
If an incompatible baud rate is selected the Analyser will not respond. Please disconnect and
reconnect the Analyser to return the Analyser to the default 57600 bps. A correct baud rate
selected will remain stored in memory even after power off.
The command getbaud will return the set baud rate. It can also be seen on the comment
section of the Feasa terminal window.
Test times are improved by increasing the baudrate from the default to higher values and this
can be further enhanced by decreasing the latency (windows default 16) to latency of 1. Refer
to the App Note in the documentation folder of the CD to carry out this.
Example:
To change the baud rate to 19,200 transmit the command setbaud19200 to the Analyser .
setbaud19200
OK
Back to Index
53

Table of Contents

Questions and Answers:

Feasa 10-F Specifications

General IconGeneral
BrandFeasa
Model10-F
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals