FEASA LED ANALYSER
FUNCTIONAL VERSION
Get Data Mode
getPHOTOPIC - Get which Photopic Mode is active
Transmit Receive
getphotopic 0 or 1 or 2
Where: 0 is Photopic Response OFF
1 is Response Photometric
2 is Response Radiometric
Description
The Led Analyser is shipped from the factory in photopic mode 1. This corrects the Absolute
Intensity to reproduce the photometric response (for lm, mcd). This can be verified by typing
the command getphotopic with the answer 1.
The photopic mode can be set to a radiometric response by sending the command
setphotopic2. This corrects the Absolute Intensity to reproduce the radiometric response (for
mW)
This command is only available on Software Rev F210 upwards