FEASA LED ANALYSER
FUNCTIONAL VERSION
User Calibration Mode
getYOFFSET## - Return the y Chromaticity offset
Transmit Receive
getyoffset## ±0.yyy
Where:
## represents the Fiber Number and is a number in the range 01 – 20.
±0.yyy represents the y Chromaticity offset
Description
This command is used to return the y Chromaticity offset for the LED under the Fiber number
##. The value of this offset must be set by the setyoffset## command. The default value
is 0.000.
Example:
The PC transmits getyoffset01 to the LED Analyser to instruct it to send the stored y
Chromaticity offset for the LED positioned under Fiber No 1. The LED Analyser will return a
string ±0.yyy to the PC.
getyoffset01
-0.025