FEASA LED ANALYSER
FUNCTIONAL VERSION
Sequence Capture Mode
Set/Put SequenceWait##(#) - Set Sequence Wait Time
Transmit Receive
setsequencewait##
putsequencewait##
setsequencewait###
putsequencewait###
OK
OK
OK
OK
Where:
## represents the wait time in milliseconds (00 – 99 msec)
### represents the wait time in milliseconds (000 – 999 msec)
Description
This command is used to set the Wait time between captures. The Wait time can be set to any
value in the range 000 – 999 milliseconds. This Wait time is set for all Fiber Channels enabled
by the setSequencemode# or putSequencemode# command.
PutSequencewait##, PutSequencewait### commands are stored in RAM memory while
SetSequencewait##, SetSequencewait### commands are stored in Flash memory.
Example:
The PC transmits putsequencewait10 to the LED Analyser to instruct it to set the wait time
between captures to 10 msec. The LED Analyser will return a string OK to the PC.
putsequencewait10
OK
To determine what Wait Time between Sequence Captures is set to, use the command
getsequencewait### . It will return a time in milliseconds.
getsequencewait
005 ms (5 milliseconds)
Back to Index