ActionStep
Empty and discard the remaining tubes:4
Press the QuickRelease device to the lock position, and empty the remaining tubes.
Pull the QuickRelease device to the release position, and discard the tubes.
Discard the tubesEmpty the tubes
Fraction collector tubes and bottles
The tubes and bottles used in Fraction collector F9-C must fulfill the requirements listed in the table
below. Examples of manufacturers are also listed in the table.
Examples of manufac-
turers
Height (mm)Diameter (mm)Tube or bottle
size (ml)
Max.Min.Max.Min.
NUNC™565011.510.53
VWR™76701210.55
BD™ Biosciences, VWR1029613.3128
BD Biosciences120114171615
BD Biosciences116110302850
Nalgene™, Kautex™121-L: 64.5
W: 64
1
L: 55
W: 55
1
250 ml bottle
1
Length and width of the rectangular bottle base
100 ÄKTA pure User Manual 29119969 AB
3 ÄKTA pure external modules
3.2 Fraction collector F9-C
3.2.3 Cassettes, Cassette tray and racks