Maximum flow rate
Fraction collection can be performed at different maximum flow rates depending on the size of the
tubes that are used. The table below lists the maximum flow rates for the Fraction collector tubes.
Maximum flow rate [ml/min]Tube size [ml]
153
258
5015
15050
150250
Deep well plates
Requirements
The deep well plates used in Fraction collector F9-C must fulfill the requirements listed in the table
below.
SpecificationProperty
24, 48, or 96No. of wells
Square, not cylindricalShape of wells
10, 5, or 2 mlWell volume
Approved deep well plates
The plates listed in the table below are tested and approved by GE to be used with Fraction collector
F9-C.
Part no.ManufacturerPlate type
7701-5200 (Whatman™)GE96 deep well plate
353966BD Biosciences
780270Greiner Bio-One
219009Porvair Sciences
S30009Seahorse Bioscience™
951033405/ 0030 501.306Eppendorf™
7701-5500 (Whatman)GE48 deep well plate
S30004Seahorse Bioscience
ÄKTA pure User Manual 29119969 AB 101
3 ÄKTA pure external modules
3.2 Fraction collector F9-C
3.2.3 Cassettes, Cassette tray and racks

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AKTApure and is the answer not in the manual?

GE AKTApure Specifications

General IconGeneral
BrandGE
ModelAKTApure
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals