Main Power Supply replacement
Vivid E80/E90/E95 – Service Manual 8-105
GC091052 Rev. 3
Main Power Supply replacement
Preparations
When preparing for the replacement, you must perform the
following steps:
1. Power down the Ultrasound system.
2. Disconnect the Mains Power Cable from the wall outlet.
3. Disconnect all probes and external I/O cabling.
4. Remove both Side Cover.
5. Remove the Top Cover.
6. Remove the Filter Cover.
7. Remove the Rear Cover.
Main Power Supply removal procedure
Follow these steps to remove the Main Power Supply:
1. Disconnect the cables on top of the power supply.
2. Remove the three screws that secure the Main Power
Supply.
3. Ease the Main Power Supply away from the fang bracket.
4. Pull/slide the entire Main Power Supply backwards until the
connectors to the Front End Processor are unseated.
5. Lift the Main Power Supply away.
Main Power Supply installation procedure
Follow these steps to install the Main Power Supply:
1. Place the Main Power Supply base on the Ultrasound
systems frame, and then tilt the top of the Main Power
Supply toward the rear of the Card Cage until the Main
Power Supply is vertical.
2. Slide the Main Power Supply forward to seat the connectors
on the Card Cage.
3. Engage the tab onto the Fang.
4. Install the three screws.

Table of Contents

Other manuals for GE Vivid E90

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E90 and is the answer not in the manual?

GE Vivid E90 Specifications

General IconGeneral
BrandGE
ModelVivid E90
CategoryMedical Equipment
LanguageEnglish

Related product manuals