General Ultrasound system requirements
Vivid E80/E90/E95 – Service Manual 2-9
GC091052 Rev. 3
Probes environmental requirements
Operation and storage temperatures for probes
Table 2-5: Operation and storage temperatures for probes
Conditions Electronic PAMPTE
Operation: 10 to 40 ºC (50 to 104 ºF) 5 to 42.7 ºC (41 to 108,9 ºF)
Storage: -20 to 50 ºC (-4 to 122 ºF) -20 to 60 ºC (-4 to 140 ºF)
Temperature in degrees Celsius (ºC) conversion to degrees F (ºF):
(ºF) = (ºC * 9/5) + 32
CAUTION
PAMPTE probes are designed for storage temperatures of -20
to +60 degrees C (-4 to +140 degrees F).
Electronic probes are designed for storage temperatures of -20
to +50 degrees C (-4 to +122 degrees F).
When exposed to large temperature variations, the product
should be kept at room temperature the needed time to
stabilize its temperature before use.